INPUT SPECIFICATION
Input Voltage: 90-264Vac typical.
Input Frequency: 47-63 Hz. (Nominal 50/60Hz.).
Input Current: 0.16A @115Vac./0.1A @230Vac typical.
Inrush Current: 4.2Arms@230Vac typical.
Input Fuse: Use external fuse.
Dielectric Withstand: Meet IEC60601-1. 4,000Vac-Output/Input. 2,000Vac-Input/GND. 1,000Vac-Output/GND.
EMI: Meet CISPR PUB. 22/FCC Class B.
Hold-up time: Typ. 22.4 mS @115Vac, 130mS @230Vac.
Green Power: 0.25W@132Vac, 0.51W@230Vac for single.
Earth Leakage: Less than 0.16mA @230Vac.
Remote ON/OFF: ON(Enable)=Open. OFF(Disable)=Short.
|
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Wattage: 10 Watts typical.
Line Regulation: Various with output voltage. Typ.±0.1% .
Load Regulation: Various with output voltage. Main VO1 ± 0.1-2.0% typical. Aux.VO2 ± 0.7-1.2% typical. Aux.VO3 ± 0.1% typical.
Total Regulation: 0.1-1.9% typical.
Noise & ripple: 1.0% typical peak to peak. Typ. 100mV for 3.3V/5V.
OVP: Built-in on main output VO1.
Adjustability: Available at main output VO1.
Overload Protection (OLP): Fully protected against output overload and short circuit. Typical 125-150% of rating output load. Con2sult the factory for special OLP setting.
|