UNIVERSAL INPUT HARMONIC CORRECTION AC-DC MEDICAL AND IEC APPLICATION OPEN FRAME SINGLE & MULTI-OUTPUT 150 WATTS INTERNAL SWITCHING POWER SUPPLIES HIMP151 SERIES
|
FEATURES:
5” X 3”* X 1.38” COMPACT SIZE
ACCOMMODATE UNIVERSAL AC SOURCES
MEET IEC 61000-3-2 HARMONIC CORRECTION
MEET UNIVERSAL SAFETY STANDARDS
EMI MEET CISPR EN 55022/EN 55011 & FCC CLASS B CONDUCTED
CE MARKING COMPLIANCE |
FEATURES |
INPUT SPECIFICATION
Input Voltage: Typ. 90-264Vac with PFC.
Input Connector: V-M connector.
Input Frequency: 47-63Hz.
Inrush Current: 9.3-11.3Arms @230Vac.
Input Current: Typ. 2.2A @115Vac, 0.9A @230Vac.
Dielectric Withstand: Meet IEC 60950-1.
EMI: Meet CISPR EN 55022 & FCC Class B conducted.
Meet CISPR EN 55011 & FCC Class B conducted.
Hold-up Time: Typ. 90mS @115Vac & 230Vac.
Power Factor & Harmonic Correction: Meet IEC 61000-3-2,
PF typ. 0.99 @ full load.
Over Temperature Protection: Installed.
Leakage Current: 0.40-0.50mA @230Vac touch current. |
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Connector: V-M connector.
Output Wattage: Max. 150 watts.
Line Regulation: Typ. 1.0%.
Load Regulation: Single O/P VO1 ±3.0%. Dual O/P VO1 ±3.0%
VO2 ±1.0% Triple O/P VO1 ± 3.0% (± 8% for 3.3V),
VO2/VO3 ±3.0%. Quad O/P VO1/VO2/VO3/VO4 ±3.0%
Noise & Ripple: Typ. 1.0-2.0% peak to peak.
OVP: Built-in at VO1 and VO2.
Adjustability: Available at VO1 and VO2.
Overload Protection
(OLP): Fully protected against output overload and short circuit.
OLP set at about 125-150% rating output wattage. Consult the factory
for OLP setting. |
GENERAL SPECIFICATION |
|
Efficiency: Typ. 76-80% various with output.
Switching Frequency: 68K Hz.
Circuit Topology: Single Fixed frequency forward circuit.
Dual/Triple/Quad: Fixed freq. flyback circuit.
Transient Response: Output voltage returns in less
than 14.8-25.2mS following a 25% load change.
Safety Standard: IEC 60950-1 Class I for IEC use. IEC
60601-1 Class I for Medical use. |
Operating Temperature: 0 to +50°C under 30 CFM forced air cooling
for full load.
Storage Temperature: -20 to +85°C.
Temperature Coefficient: Typ. 0.04% /°C.
Cooling: Requested 30 CFM forced air flow for full
load.
Construction: U-bracket format.
Power Density: 6.52-7.24 Watts / Cubic Inch . |
OUTPUT VOLTAGE / CURRENT RATINGS CHART
SINGLE OUTPUT (under air cooling)
DUAL OUTPUT (under forced air cooling)
TRIPLE OUTPUT (under forced air cooling)
QUAD OUTPUT (under forced air cooling)
MECHANICAL DIMENSIONS: MM [INCHES] WEIGHT: 385.0g (13.4 Oz.)
INPUT & OUTPUT CONNECTORS PIN ASSIGNMENT