FEATURES:

5” X 3”* X 1.38” COMPACT SIZE
ACCOMMODATE UNIVERSAL AC SOURCES
MEET IEC 61000-3-2 HARMONIC CORRECTION
MEET UNIVERSAL SAFETY STANDARDS
EMI MEET CISPR EN 55022/EN 55011 & FCC CLASS B CONDUCTED
CE MARKING COMPLIANCE

FEATURES

INPUT SPECIFICATION

Input Voltage: Typ. 90-264Vac with PFC.
Input Connector: V-M connector.
Input Frequency: 47-63Hz.
Inrush Current: 9.3-11.3Arms @230Vac.
Input Current: Typ. 2.2A @115Vac, 0.9A @230Vac.
Dielectric Withstand: Meet IEC 60950-1.
EMI: Meet CISPR EN 55022 & FCC Class B conducted. Meet CISPR EN 55011 & FCC Class B conducted.
Hold-up Time: Typ. 90mS @115Vac & 230Vac.
Power Factor & Harmonic Correction: Meet IEC 61000-3-2, PF typ. 0.99 @ full load.
Over Temperature Protection: Installed.
Leakage Current: 0.40-0.50mA @230Vac touch current.

OUTPUT SPECIFICATION

Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Connector: V-M connector.
Output Wattage: Max. 150 watts.
Line Regulation: Typ. 1.0%.
Load Regulation: Single O/P VO1 ±3.0%. Dual O/P VO1 ±3.0% VO2 ±1.0% Triple O/P VO1 ± 3.0% (± 8% for 3.3V), VO2/VO3 ±3.0%. Quad O/P VO1/VO2/VO3/VO4 ±3.0%
Noise & Ripple: Typ. 1.0-2.0% peak to peak.
OVP: Built-in at VO1 and VO2.
Adjustability: Available at VO1 and VO2.
Overload Protection (OLP): Fully protected against output overload and short circuit. OLP set at about 125-150% rating output wattage. Consult the factory for OLP setting.

GENERAL SPECIFICATION

Efficiency: Typ. 76-80% various with output.
Switching Frequency: 68K Hz.
Circuit Topology: Single Fixed frequency forward circuit.
Dual/Triple/Quad: Fixed freq. flyback circuit.
Transient Response: Output voltage returns in less than 14.8-25.2mS following a 25% load change.
Safety Standard: IEC 60950-1 Class I for IEC use. IEC 60601-1 Class I for Medical use.

Operating Temperature: 0 to +50°C under 30 CFM forced air cooling for full load.
Storage Temperature: -20 to +85°C.
Temperature Coefficient: Typ. 0.04% /°C.
Cooling: Requested 30 CFM forced air flow for full load.
Construction: U-bracket format.
Power Density: 6.52-7.24 Watts / Cubic Inch .

OUTPUT VOLTAGE / CURRENT RATINGS CHART

SINGLE OUTPUT (under air cooling)

SINGLE OUTPUT

DUAL OUTPUT (under forced air cooling)

DUAL OUTPUT

TRIPLE OUTPUT (under forced air cooling)

TRIPLE OUTPUT

QUAD OUTPUT (under forced air cooling)

QUAD OUTPUT

MECHANICAL DIMENSIONS: MM [INCHES] WEIGHT: 385.0g (13.4 Oz.)

MECHANICAL DIMENSIONS: MM [INCHES] WEIGHT: 240g (8.11 Oz.)

INPUT & OUTPUT CONNECTORS PIN ASSIGNMENT

INPUT & OUTPUT CONNECTORS PIN ASSIGNMENT